40W Single Output Class II Medical Power Supplies
BSM40
series
WWW.ASTRODYNE.COM
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
INPUT SPECIFICATIONS
Input Voltage Range
Frequency Range
Input Current (90 / 264 Vin)
Inrush Current (115 / 230 Vin)
Leakage Current (264V / 50Hz)
90-264 VAC
47-63 Hz
1A / 0.4A
30A / 60A typical
*
<0.25mA
OUTPUT SPECIFICATIONS
Voltage and Current (Note 3)
Turn On Delay Time
Rise Time
Load Regulation (20%-FL)
Line Regulation (LL-HL)
Preset Accuracy (FL, 115Vin)
Transient Response
Over-Shoot and Under-Shoot
Ripple/Noise (Note 1 & 4)
Over Voltage Protection
Current Limit, Self Recovering
Short Circuit Protection
Hold Up Time (115V / 60Hz)
Capacitive Loading (Note 5)
See Selection Chart
<1S
<30mS
± 1% max
± 1% max
1%
See Page 3
<10% of O/P Voltage
See Selection Chart
Latching re-power
150% max., hiccup mode
*
Latching, Auto Recover
*
8 mS, typ.
See Selection Chart
PHYSICAL SPECIFICATIONS
Size (Inches / mm)
Construction
Weight
2” x 4” x 1.17” / 50.8 x 101.6 x 29.6
Open Frame
5.61oz (157.2g)
GENERAL SPECIFICATIONS
Input-Out Isolation
Insulation Resistance
Efficiency (typ.)
Switching Frequency
Safety
4000VAC
≥20M; 500VDC, 1S I/P-O/P
See Selection Chart
65Khz, (fixed, typical)
UL/cUL: UL60601-1, Class II
TUV: EN60601-1, Class II
CE: EN60601-1-2, Class II
CB: IEC60601-1, Class II
CSA: C22.2 60601.1, Class II
ENVIRONMENTAL SPECIFICATIONS
Oper. Temperature (Note 2)
Storage Temperature
Relative Humidity
ESD
RS
EFT
Surge
CS
DIPS
EMI
MTBF
Vibration
Drop Test
-10 to +70°C, See Derate
-25 to +85°C
*
0-95%
*
IEC61000-4-2
IEC61000-4-3
IEC61000-4-4
IEC61000-4-5
IEC61000-4-6
IEC61000-4-11
EN55022B / CISPR 11 B
196,000 Hrs
Mil Std 217, 25°C
2G Peak, 10-500Hz, 3 Axis, 30 min
70 cm Height
NOTES
1. Make all measurements directly at the pins of the supply
2. Specified for free air convection cooling
3. 115Vin minimum required for full load start up
4. Measured by paralleling 47uF/EC and 0.1uF ceramic capacitors on
the output at a 20MHz band-width
5. The power supply should start up and operate normally into these
capacitive loads within specified input voltage and output current ranges
over the specified operating temperature range and according to Derate.
*
These are stress ratings. Exposure of the devices to any of these conditions
may adversely affect long term reliability. Proper operation under conditions
other than the standard operating conditions is neither warranteed nor implied.
All specifications are typical at nominal input, full load, and 25°C unless other-
wise noted
Astrodyne products are not authorized or warranteed for use
as critical components in life support systems, equipment used
in hazardous environments, nuclear controls systems, or other
mission-critical applications.