CCD area image sensor
S7017 series
KMPDC0087EA
G
Line binning
Parameter
Symbol
Tpwv
Tprv, Tpfv
Tpwh
Tprh, Tpfh
-
Tpws
Tprs, Tpfs
-
Tpwr
Tprr, Tpfr
Tovr
Remark
Min.
6 *
19
200
250
10
-
250
10
-
100
5
3
Typ.
-
-
-
-
50
-
-
50
-
-
-
Max.
-
-
-
-
-
-
-
-
-
-
-
Unit
μs
ns
ns
ns
%
ns
ns
%
ns
ns
μs
Pulse width
Rise and fall time
Pulse width
Rise and fall time
Duty ratio
Pulse width
Rise and fall time
Duty ratio
Pulse width
Rise and fall time
Overlap time
P1V, P2V, TG
*
18
P1H, P2H
*
18
SG
-
RG
-
TG – P1H
-
*18: Symmetrical pulses should be overlapped at 50 % of maximum amplitude.
*19: In case of S7017-1007.
I
Specifications of built-in TE-cooler
Parameter
Internal resistance
Symbol
Rint
Condition
Min.
-
Typ.
1.6
Max.
-
Unit
Ta=27 °C
Th *
21
=27 °C
T *
22
=
Tmax
Th*
21
=27 °C
T=
Tmax
I=Imax
Tc *
24
=Th *
21
=27 °C
I=Imax
Maximum current *
20
Imax
-
-
4.4
A
Maximum voltage
Vmax
-
-
7.4
V
Maximum heat absorption *
23
Qmax
-
-
3.0
W
Maximum temperature at hot side
CCD temperature
*20: If the current is greater than Imax, the heat absorption begins to decrease due to the Joule heat. It should be noted that
this value is not a damage threshold. To protect the thermoelectric cooler (Peltier element) and maintain stable operation,
the supply current should be less than 60 % of this maximum current.
*21: Temperature at hot side of thermoelectric cooler.
*22:
T=Th - Tc
*23: This is a theoretical heat absorption level that offsets the temperature difference in the Peltier element when the maximum
current is supplied to the unit.
*24: Temperature at cool side of thermoelectric cooler.
-
-
-
-
-
50
-50
°C
°C
Ta=25 °C
-70
INTEGRATION PERIOD
(Shutter must be open)
VERTICAL BINNING PERIOD
(Shutter must be closed)
3..126
3..254
127
255
128
←
256
←
P1V
P2V, TG
P1H
P2H, SG
READOUT PERIOD (Shutter must be closed)
124+4 (ISOLATION): S7017-1007
252+4 (ISOLATION): S7017-1008
Tpwv
Tovr
Tpwh, Tpws
1
Tpwr
2
3
1043
1044
4..1042
1
2
D19
D2
D3..D10, S1..S1024, D11..D18
D1
D20
RG
OS
7