Mini-spectrometer TG series
C9404MC, C9405MC
2
I
Specifications
TG-UV-MOS
C9404MC
TG-SWNIR-MOS
C9405MC
Parameter
Unit
Number of pixels
Spectral response range
Spectral resolution Max.
(Spectral response half width) *
2
Wavelength reproducibility
*
4
Wavelength temperature dependence
Spectral stray light *
2,
*
5
Broadband stray light *
2,
*
6
Slit *
7
A/D conversion
Integration time
Optical NA *
8
512
pixels
nm
200 to 400
500 to 1100
3
5
*
3
nm
±0.1
±0.2
nm
0.02
nm/°C
dB
dB
μm
bit
ms
-
-35
-27
-35
-27
140 (H) × 500 (V)
70 (H) × 2500 (V)
16
5 to 10000
0.11
0.22
Image sensor
CMOS linear image sensor
(S8378-512Q)
NMOS linear image sensor
(S8381-512Q)
SMA905D
USB1.1
-
Connector for optical fiber
Interface
Current consumption *
9
Operating temperature *
10
Storage temperature
Dimensions
*2: Depends on the slit opening. Values were measured with the slit opening listed in the table.
*3:
λ
=550 to 1100 nm
*4: Measured under constant light input conditions.
*5: When monochromatic light of the following wavelengths is input, spectral stray light is defined as the ratio of the count
measured at the input wavelength, to the count measured in a region of the input wavelength ±20 nm (C9404MC) or ±40
nm (C9405MC).
C9404MC: 300 nm, C9405MC: 800 nm
*6: This is the ratio of the transmittance measured with light passing through the following optical filters to the transmittance
measured in the blocking region.
C9404MC: WG320, C9405MC: RG850
*7: Entrance slit aperture size
*8: Numerical aperture (solid angle)
*9: USB bus power
*10: No condensation
I
Optical component layout
TG series mini-spectrometers use a transmission holographic
grating made of quartz and precision optical components
arranged on a rugged optical base, making it possible to
deliver high throughput and highly accurate optical charac-
teristics.
-
-
100
100
mA
°C
°C
mm
+5 to +40
-20 to +70
40 (W) × 106 (D) × 86 (H)
ENTRANCE SLIT
FOCUSING LENS
IMAGE SENSOR
COLLIMATING LENS
TRANSMISSION GRATING
QUARTZ CELL (For holding liquid sample)
MINI-SPECTROMETER
USB CABLE
PC
UV-VIS FIBER LIGHT SOURCE
(Deuterium lamp and Halogen lamp)
L10290
FIBER
KACCC0256EA
I
Connection example
(transmission light measurement)
Light to be measured is guided into the entrance port of TG
series through an optical fiber and the spectrum measured
with the built-in image sensor is output through the USB port
to a PC for data acquisition. There are no moving parts inside
the unit so stable measurements are obtained at all times. An
optical fiber that guides light input from external sources al-
lows a flexible measurement setup.
KACCC0257EB