Turn key to "SET" or "LOCK"
position.
Key is on “OFF” position
One or more emitter cassettes are
not fully inserted.
Push in emitter cassette(s)
completely.
No Power
Check power cord or fuse.
Key is in "Lock " Position
(Blue indicator On).
Turn key to “Set” position
Button depressed too quickly,
or pressed too lightly.
Push and hold buttons firmly
"PWR" light is
flashing red.
Emitter pins are dirty or
contaminated.
Remove emitter cassette
and clean emitters.
"PWR" light is
flashing after emitter
cassettes have been
cleaned.
Refer to page 4 for
calibration instructions.
There is no audible
alarm when "PWR"
light is flashing red.
PROBLEM
POSSIBLE CAUSE
SOLUTION
Unit won't turn on
Fan speed or
Balance won't
adjust
Unit is out of calibration.
Alarm switch/jumper is set
to off.
Refer to page 4 for
switch/jumper location.
Limited Warranty
Charleswater Europe expressly warrants that for a period of five (5) years from the date of purchase Charleswater Europe CE Ionizers will be free of defects in material
(parts) and workmanship (labour). Within the warranty period, a credit for purchase of replacement Charleswater Europe CE Ionizers, or, at our option, the CE Ionizer
will be repaired or replaced free of charge. If product credit is issued, the amount will be calculated by multiplying the unused portion of the expected five year life
times the original unit purchase price. Call Customer Service at 00 44 (0) 1892-665313 for Return Material Authorisation (RMA) and proper shipping instructions and
address. Any unit under warranty should be shipped prepaid to the Charleswater Europe factory. You should include a copy of your original packing slip, invoice, or
other proof of date of purchase. Warranty replacements will take approximately two weeks.
If your unit is out of warranty, Charleswater Europe will quote repair charges necessary to bring your unit up to factory standards. Call Customer Service at
00 44 (0) 1892-665313 for a Return Material Authorisation (RMA) proper shipping instructions and address. Units should be shipped prepaid.
Warranty Exclusions
THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL OTHER PRODUCT WARRANTIES, EXPRESSED AND IMPLIED, INCLUDING MERCHANT-
ABILITY AND FITNESS FOR A PARTICULAR PURPOSE WHICH ARE SPECIFICALLY DISCLAIMED. The express warranty will not apply to defects or damage due
to accidents, neglect, misuse, alterations, operator error, or failure to properly maintain, clean or repair products.
Limit of Liability
Electronic ionisers use high voltage corona discharge and should not be used in or near flammable or explosive environments. In no event will Charleswater Europe or
any seller be responsible or liable for any injury, loss or damage, direct or consequential, arising out of the use of or the inability to use the
product. Before using, users
shall determine the suitability
of the product for their intended use, and users assume all risk and liability whatsoever in connection therewith.
PPE-5060.E
Page 6 of 6
Charleswater Europe Ltd.
Unit 17. Millbrook Business Park, Sybron Way Crowborough, East Sussex TN6 3JZ United Kingdom
Phone: 00 44 (0) 1892-665313, Fax: 00 44 (0) 1892-668838 E-mail: info@charleswater.co.uk, Internet: www.charleswater.co.uk